Home > Products >Trauma and Extremities >Clavicular Hook LCP (L&R)
Clavicular Hook LCP (L&R)
Overview

Part Number: 1310-0735 ~ 1310-1635 (Left F1295) 6310-0735 ~ 6310-1635 (Left F138)
Part Number: 1311-0735 ~ 1311-1635 (Right F1295) 6311-0735 ~ 6311-1635 (Right F138)


Clavicular Hook LCP (Left & Right)

The SurgTech Trauma System is a plate and screw fixation system. Plates and screws are offered in “mini”,“small” and “large” set sizes in a variety of shapes based upon the anatomical fixation required.

TC20: New Material Without Vanadium Trouble

The LCP and instrument sets are designed and manufactured exclusively by SurgTech Inc.

Features
The SurgTech Trauma System is available in a wide range of sizes: 74-162mm L / 4,5,7,9,11 hole / 3.6mm t / 10.8mm w.
The features that deserve special attention are:
Features
  • Matching Screws
  • Low Profile
  • The Groove Reduces The Contact
  • More Specifications
  • Anatomic Pre-shaping & Anatomic Design
    Low profile design to reduce irritation of soft tissues.
    The groove reduces the contact area with periosteum. Limiting this contact will reduce damage to the periosteum.
    More specifications, more choices
    Anatomic pre-shaping & anatomic design, 12° pre-bending on the left and right sides, easy to place.

 

Materials: TC20     Full Name: Ti6Al7Nb
It is a new type of implant material with high technology content and high economic value. Standards implemented in different countries and regions include: ISO5832-11, ASTM F1295, GB/T13810, GB 23102.

Chemical Composition

 

TC20 VS TC4


Differences between TC20 and TC4
TC20 (Ti6Al7Nb) and TC4(Ti-6Al-4V) belong to alpha + beta-titanium alloy. The main difference is that TC20 uses the non-toxic element niobium (Nb) instead of the toxic element vanadium (V) in TC4.

 

Advantage of Difference

1. The potential cytotoxicity of element V is eliminated and the material has good cytocompatibility.

2. Nb element can promote the density increase of passivation film formed by surface oxidation of ti-6al-7nb alloy, so that the material has good corrosion resistance.

3. The oxides and hydroxides of titanium, aluminum and niobium have reached saturation, and the dissolved metals in the implants will not be ionized or transferred in the body, making them have good biocompatibility.

Instruments
Instrument Set include:
Instruments
  • A Full Set of Tools Layer I
  • A Full Set of Tools Layer II
    1. Drill φ2.8x165; 2. Probe φ2x160; 3. Broken Nail Extractor; 4. Countersink; 5. Skid Extractor; 6. Drill φ2.5; 7. Plum Blossom Screwdriver; 8. Bender (right); 9. T-type Quick Change Handle; 10. Depth Gauge; 11. Drill Bush  φ2.8; 12. Screw Tap 3.5; 13. Screw Tap 4.0; 14. Thread Guide Sleeve; 15. Guide Handle; 16. Drill Bushing φ2.5; 17. Guide Sleeve φ3.5; 18. Guide Sleeve φ4.0; 19. Eccentric Drill Sleeve φ2.5; 20. Wrench; 21. Bender (left); 22. Drill Bushing φ2.5; 23. Drill φ3.5; 24. Nail Holder; 25. Stopper φ3.0; 26. Fast Switching Adapter.
    27. Torque Wrench; 28. Tip Reduction Forceps (Middle); 29. Bending Reset Forceps; 30. Centralized Bone Holding Forceps (Middle); 31. Ratchet Wrench.
Indication

Lateral Clavicle Fracture and Acromioclavicular Dislocation.


Follow us:

Copyright © 2019 SurgTech Inc. All Rights Reserved. Designed by Wanhu